(Click to view full image)

Project Overview

Client Type

Leading Tier-1 Medical Imaging OEM

Production Volume

15,000+ units/year

Material Used

Unfilled Polyetherimide (PEI)

8 weeks from DFM to PPAP

8 weeks from DFM to PPAP

Executive Summary

  • Client: Leading tier-1 medical device manufacturer in China.
  • Application: RF coil balun (cable trap) enclosures & housings for 1.5T and 3.0T MRI systems.
  • Core Challenge: Establishing a highly conductive, geometrically stable EMI RFI shielding layer over a plastic substrate. This layer suppresses common-mode currents to prevent image artifacts and patient RF burns.
  • Material Specification: Unfilled polyetherimide (PEI). This PEI material delivers inherent UL94 V-0 flame retardancy, low outgassing, and critical dielectric stability under thermal stress. We omitted glass fillers to maximize surface adhesion for metallization.
  • Manufacturing Process: Tight-tolerance medical injection molding services paired with high-adhesion thick copper plating services. The copper layer thickness specifically exceeds RF skin depth requirements at designated MRI operating frequencies.
kravzik-rf-cable-trap-assembly-details
kravzik-rf-cable-trap-assembly-details

The Engineering Challenges

The Electrical Challenge: RF Shielding & Patient Safety

MRI coils transmit signals via coaxial cables. If stray common-mode currents travel along the cable shield, they degrade image quality. This causes visible image artifacts and generates localized heat. The balun must act as a precise high-impedance choke. It blocks these unwanted currents and prevents severe patient RF burns. To achieve this, the component requires flawless surface conductivity. We rely on uniform emi rfi shielding to deliver the necessary electrical performance.

The Mechanical Challenge: Dimensional Stability Under Thermal Stress

MRI scanners generate continuous heat during normal operation. The balun housing clamps directly onto the cable to form a strictly calibrated resonant circuit. If the plastic substrate warps due to thermal expansion, the mechanical grip fails. Any physical deformation will instantly detune the RF circuit. The molded pei material must maintain absolute dimensional stability. We must lock all physical dimensions within tight tolerances, even under sustained thermal stress.

kravzik-medical-grade-rf-shielding-component
kravzik-medical-grade-rf-shielding-component

The Chemical Challenge: Copper Plating Adhesion on Inert Plastics

Bonding a thick layer of conductive copper to chemically resistant plastics is notoriously difficult. Our goal is to achieve massive adhesion strength. The metallized layer from our copper plating services must be entirely uniform. It cannot peel, flake, or blister under any circumstances. The component will endure repeated thermal cycling and mechanical stress in the field. The copper-to-plastic bond must remain permanent and electrically continuous.

Material Selection Strategy

Inherent Flame Retardancy (UL94 V-0)

Medical environments demand absolute fire safety. Our material analysis led us to specify unfilled pei material for the balun substrate. PEI delivers inherent UL94 V-0 flame retardancy. It achieves this strict rating without relying on halogenated chemical additives. This hard, highly rigid polymer withstands rigorous clinical safety protocols. It also maintains exceptionally low smoke toxicity under thermal fault conditions.

kravzik-high-conductivity-copper-plating
kravzik-high-conductivity-copper-plating

Dielectric Stability in RF Environments

The plastic substrate must remain electrically transparent. It cannot interfere with sensitive MRI RF signals. PEI maintains a highly stable dielectric constant (Dk). It also features a low dissipation factor (Df) across broad frequency and temperature ranges. This dielectric stability is crucial. It ensures the physical housing never detunes the calibrated resonant circuit during active scanning operations.

The “Unfilled” Imperative for Plating Adhesion

We deliberately selected an unfilled resin. This represents a critical engineering tradeoff. Product designers often add glass fibers (GF) to plastics to increase tensile strength. However, glass fibers create uncontrolled surface micro-roughness. During molding, exposed fiber ends migrate to the plastic surface. When we apply thick copper plating services, these microscopic imperfections trap active chemicals. This causes severe plating blistering, poor adhesion, and uneven RF skin depth. Unfilled PEI provides a flawless, glass-smooth substrate. This maximizes the chemical copper bond strength while retaining ample baseline rigidity for the mechanical clamping function.

kravzik-mri-balun-parting-line-accuracy
kravzik-mri-balun-parting-line-accuracy

Manufacturing Process Control

Precision Tooling for RF Integrity

Balun enclosures & housings typically use a split-shell design that clamps around the coaxial cable. Managing the inherent mold shrinkage of PEI is critical. Any warping or mismatch at the parting line creates microscopic physical gaps. These gaps immediately cause RF leakage. Our injection mold design & making team strictly controls mold temperatures and cavity pressures. This ensures the two plastic halves mate flawlessly to form an unbroken shield.

Stress-Free PEI Injection Molding

PEI requires exceptionally high melt temperatures during processing. If operators inject the resin too quickly, severe internal stresses lock inside the plastic matrix. When this stressed part enters the aggressive chemical etching bath for plating, it suffers from micro-cracking. We utilize precision medical injection molding services to calibrate melt temperatures and injection velocities. This specific process window yields a structurally relaxed, stress-free substrate. It is perfectly conditioned for permanent chemical bonding.

kravzik-balun-assembly-in-hand-scale
kravzik-balun-assembly-in-hand-scale

High-Adhesion Copper Plating

Our surface treatment relies on a strict multi-stage chemical sequence. First, we chemically roughen the bare pei material to create microscopic anchor points. Next, we deposit an initial electroless copper seed layer for baseline conductivity. We finish with a robust electrolytic copper build-up via our copper plating services. We engineer the final plating thickness to comfortably exceed the RF skin depth required at specific MRI operating frequencies. Before shipment, every batch undergoes a rigorous cross-hatch test (ASTM D3359). This verifies absolute adhesion strength and guarantees the shield will not flake.

kravzik-mri-balun-quality-verification
kravzik-mri-balun-quality-verification

Quality Assurance & Compliance

Dimensional Metrology & Cpk Tracking

The mating halves of the balun require absolute physical alignment to prevent RF leakage. We use full-scale Coordinate Measuring Machine (CMM) inspection for all first-article approvals. During mass production, we implement continuous Statistical Process Control (SPC). We maintain a strict target Cpk > 1.33 for all critical clamping dimensions to ensure structural integrity.

Plating Verification & Thermal Cycling

The copper layer must survive continuous heat generated by the active MRI coil. We utilize X-ray Fluorescence (XRF) equipment to map copper thickness across complex part geometries. This guarantees the metallization reliably exceeds the required RF skin depth. We also subject batch samples to harsh thermal cycling stress tests. This proves the surface will never delaminate or blister under field conditions.

kravzik-copper-plated-plastic-mri-component
kravzik-copper-plated-plastic-mri-component

Medical Industry Compliance

We mandate strict batch traceability for all raw materials and chemical baths. The final plated enclosures & housings are entirely RoHS and REACH compliant. This fulfills the strict environmental and material safety baseline demanded by global tier-1 medical equipment manufacturers.

Conclusion & Technical Capabilities

Achieving high-adhesion, dimensionally stable copper plating on unfilled pei material sets a strict manufacturing benchmark. We established this proven process framework on demanding MRI balun assemblies. This exact methodology translates directly to processing other advanced engineering plastics. We apply these strict control parameters to molding and metallizing PEEK and PPS for the medical sector.

Our technical footprint extends well beyond MRI components. This specific combination of tight-tolerance molding and robust metal-on-plastic metallization forms our production baseline. We utilize this baseline to manufacture highly demanding emi rfi shielding components. Typical applications include surgical robotics frameworks, high-frequency diagnostic equipment, and ruggedized sensor enclosures & housings.

We rely on strict data to evaluate your project feasibility. Submit your 3D CAD geometry (STEP files) or 2D technical drawings to our engineering team. We will execute a comprehensive DFM evaluation and strict tolerance assessment for your custom components.

You May Like Also

Access Our Blogs Center
  • A supplier who plates your terminals but skips the JESD201 post-plating anneal will pass the incoming dimensional check. You discover tin whiskers only after field returns arrive. This checklist audits plating specification, pre-treatment protocols, and post-plating verification. Go through every item.

    kravzik-high-precision-stamped-electrical-terminals (8)

    A PPAP package submitted with a compliance letter but without the Cpk capability study will pass the document checklist and fail the customer SQE review. This checklist audits dimensional capability data, material cert traceability, and testing standards documentation before the submission package leaves your supplier. Go through every document.

    kravzik-high-precision-stamped-electrical-terminals (23)

    A terminal drawing that passes CAD review but lacks springback compensation data will drift out of tolerance by stroke 50,000. Your supplier catches it only at first-article rejection. This checklist audits flat blank layout, bend radii, and material grain alignment before die steel is cut. Go through every item.

    kravzik-high-precision-stamped-electrical-terminals (45)
Load More

Request DFM & Quote